Spatial accuracy of dose delivery significantly impacts the planning target volume margin in linear accelerator-based intracranial stereotactic radiosurgery

Y Yuta Takahashi R Riki Oshika R Rie Tachibana K Katsuyuki Shirai H Hiroshi Asakura M Masayoshi Miyazaki (MDX Research Center for Element Strategy, Institute of Integrated Research, Institute of Science Tokyo, Midori-ku, Yokohama 226-8503, Japan) T Tomohiro Sagawa S Shinichi Takahashi T Tsunekazu Kuwae H Hironori Kojima S Shiro Nishiyama H Hikaru Nemoto Y Yoshitomo Ishihara M Mariko Umeda K Kotaro Kijima D Daisuke Kobayashi (Department of Medical Entomology, National Institute of Infectious Diseases) K Keiji Suzuki Y Yuki Nozawa K Kento Hoshida T Tomoki Kitagawa H Hiromitsu Endo Y Yuki Matsunaga H Hiroya Itagaki M Mayumi Ishida S Shigeru Kanahara R Ryo Horita D Daisuke Hori H Hidenobu Tachibana

Abstract

Abstract The impact of three-dimensional (3D) dose delivery accuracy of C-arm linacs on the planning target volume (PTV) margin was evaluated for non-coplanar intracranial stereotactic radiosurgery (SRS). A multi-institutional 3D starshot test using beams from seven directions was conducted at 22 clinics using Varian and Elekta linacs with X-ray CT-based polymer gel dosimeters. Variability in dose delivery accuracy was observed, with the distance between the imaging isocenter and each beam exceeding 1 mm at one institution for Varian and nine institutions for Elekta. The calculated PTV margins for Varian and Elekta linacs that could cover the gross tumor volume with 95% probability at 95% of the institutions were 2.3 and 3.5 mm, respectively, in the superior–inferior direction. However, with multifactorial system management (i.e., high-accuracy 3D dose delivery with rigorous linac quality assurance, strict patient immobilization, and high intra-fractional positioning accuracy), these margins could be reduced to 1.0 mm and 1.5 mm, respectively. The findings indicate significant millimeter-level variability in 3D dose delivery accuracy among linacs installed in clinical settings. Thus, maximizing a linac’s 3D dose delivery accuracy is essential to achieve the required PTV margin in intracranial SRS.

Article Details

Volume / Issue Vol. 15, Issue 1
Published January 29, 2025
ISSN 2045-2322
Publisher Nature Portfolio

Journal Info

Scientific Reports

Nature Portfolio

ISSN: 2045-2322 Open Access Life Sciences

Authors (28)

Y

Yuta Takahashi

R

Riki Oshika

R

Rie Tachibana

K

Katsuyuki Shirai

H

Hiroshi Asakura

M

Masayoshi Miyazaki

MDX Research Center for Element Strategy, Institute of Integrated Research, Institute of Science Tokyo, Midori-ku, Yokohama 226-8503, Japan

T

Tomohiro Sagawa

S

Shinichi Takahashi

T

Tsunekazu Kuwae

H

Hironori Kojima

S

Shiro Nishiyama

H

Hikaru Nemoto

Y

Yoshitomo Ishihara

M

Mariko Umeda

K

Kotaro Kijima

D

Daisuke Kobayashi

Department of Medical Entomology, National Institute of Infectious Diseases

K

Keiji Suzuki

Y

Yuki Nozawa

K

Kento Hoshida

T

Tomoki Kitagawa

H

Hiromitsu Endo

Y

Yuki Matsunaga

H

Hiroya Itagaki

M

Mayumi Ishida

S

Shigeru Kanahara

R

Ryo Horita

D

Daisuke Hori

H

Hidenobu Tachibana