Slot fidelity and ligation-dependent tribology in personalized brackets made by casting versus selective laser melting: an in vitro study

V Viet Anh Nguyen M Minh Ngoc Tran N Ngo The Minh Pham T Thi Bich Ngoc Ha V Viet Hoang T Thi Quynh Trang Vuong

Abstract

Abstract Reliable chairside adoption of digital orthodontics depends on micrometre-scale slot fidelity and stable bracket–wire tribology. A single computer-aided design for a personalized second-premolar bracket was manufactured by lost-wax casting and by selective laser melting (SLM) (n = 36 per method). Slot height and inter-wall angle were measured on both the support-facing and non-support surfaces. Static and dynamic friction were evaluated using stainless-steel rectangular wires ligated either with tightly twisted stainless-steel ties or with elastomeric modules. Mean slot height was 480.88 ± 73.90 µm for casting and 421.47 ± 32.03 µm for SLM, against a nominal 480 µm. Overall height error did not differ between methods ( P  = 0.673), whereas angle er-ror was greater for SLM (17.76 ± 11.29°) than for casting (9.56 ± 8.88°, P  < 0.001). The support-facing wall consistently showed reduced accuracy: in casting, height error in-creased on the support side ( P  = 0.001); in SLM, both height and angle errors increased on the support side (both P  < 0.001). Across conditions, static friction exceeded dynamic friction (all P  < 0.001). With steel ligation, friction was higher for casting than SLM (median static 6.00 N vs 4.35 N, P  = 0.007; median dynamic 5.03 N vs 3.83 N, P  = 0.011). With elastomeric ligation, the ranking reversed, with higher friction for SLM (median static 2.95 N vs 2.05 N, P  < 0.001; median dynamic 2.54 N vs 2.03 N, P  = 0.003). In this standardized local in vitro model, findings suggest a surface-dependent reduction in accuracy at the support-facing wall and a fabrication-by-ligation interaction that may alter the frictional ranking between steel and elastomeric ligations. Cast brackets showed more consistent full-wire seating, whereas SLM brackets may benefit from calibrated design offsets and targeted finishing to limit undersized or tapered slots; validation beyond this setup is needed.

Article Details

Volume / Issue Vol. 16, Issue 1
Published January 06, 2026
ISSN 2045-2322
Publisher Nature Portfolio

Journal Info

Scientific Reports

Nature Portfolio

ISSN: 2045-2322 Open Access Life Sciences

Authors (6)

V

Viet Anh Nguyen

M

Minh Ngoc Tran

N

Ngo The Minh Pham

T

Thi Bich Ngoc Ha

V

Viet Hoang

T

Thi Quynh Trang Vuong