Differential sensitivity of impedance plethysmography and photoplethysmography sensors to temperature-induced peripheral vasoconstriction

S Seobin Jung S Seamus Thomson A Alexandros Pantelopoulos L Lindsey Sunden P Peter Richards S Shwetak Patel S Sam Sheng

Abstract

Abstract Impedance plethysmography (IPG) and photoplethysmography (PPG) are non-invasive techniques for measuring blood volume changes. This study investigated the differential responses of IPG and PPG to temperature-mediated vasoconstriction induced by localized cooling. Twenty-one participants underwent control and treatment conditions, with fake or real ice cubes applied to the forearm. Blood pressure remained stable, while heart rate decreased. PPG signal amplitude significantly decreased with cooling (p adj = 0.004), indicating sensitivity to superficial blood flow changes. In contrast, IPG signal amplitude remained stable (p adj = 1.0). No statistically significant differences were observed in timing-derived metrics. These findings suggest IPG is less sensitive to superficial changes in blood flow than PPG, and may be more suitable for monitoring deeper blood flow. This study provides insights into the distinct sensitivities of IPG and PPG, with implications for wearable device development and cardiovascular monitoring.

Article Details

Volume / Issue Vol. 16, Issue 1
Published January 31, 2026
ISSN 2045-2322
Publisher Nature Portfolio

Journal Info

Scientific Reports

Nature Portfolio

ISSN: 2045-2322 Open Access Life Sciences

Authors (7)

S

Seobin Jung

S

Seamus Thomson

A

Alexandros Pantelopoulos

L

Lindsey Sunden

P

Peter Richards

S

Shwetak Patel

S

Sam Sheng