Development of a Loop-Mediated Isothermal Amplification (LAMP) for the screening of Candida auris
Abstract
Background Candida auris is an emerging multidrug-resistant yeast associated with invasive infections, healthcare-associated outbreaks, and high mortality, and is often misidentified by conventional diagnostic methods. Rapid, accurate, and scalable screening tools are essential for effective infection control, particularly in high-risk settings. Materials and methods We developed a multiplex loop-mediated isothermal amplification (LAMP) assay that combines a broad-range Candida Pan target with a C. auris –specific target in a single isothermal reaction. Assay conditions were optimized for primer ratio and temperature, and analytical sensitivity was evaluated using serial dilutions of culture-derived C. albicans and C. auris DNA, as well as contrived specimens consisting of urine, swab, and whole-blood matrices. Clinical performance was assessed using 35 Candida -positive clinical specimens (blood, urine, ear swabs) and 94 non-infectious controls. Results were compared with Candida Pan qPCR and C. auris qPCR. Cross-reactivity was tested against common bacterial isolates. Results Under optimized conditions (1:1 primer ratio, 64 °C), the assay allowed species-level discrimination, with C. auris positive for both Pan and auris channels and C. albicans positive only for the Pan channel. The C. auris -specific LAMP probe detected approximately 10²–10³ cells/mL in culture-derived and contrived specimens, showing a 1–2 log improvement over C. auris qPCR (10⁴–10⁵ cells/mL), while the Pan LAMP channel detected C. auris at around 10⁵ cells/mL. In clinical specimens, Pan LAMP detected Candida spp. in 34/35 cases (97.14%) versus 32/35 (91.14%) for Pan qPCR. All C. auris –positive specimens (9/9) were detected by the multiplex LAMP assay, compared with 6/9 (66.7%) by Pan qPCR. All 94 non-infectious controls and all bacterial isolates tested negative, indicating 100% clinical specificity and absence of cross-reactivity. Conclusion The multiplex Candida Pan/ auris LAMP assay provides a rapid, highly sensitive, and specific alternative to qPCR for C. auris screening, while preserving broad Candida detection in a single isothermal reaction. Its improved analytical and clinical sensitivity suggests strong potential for use in active surveillance and infection-control programs, particularly in settings where timely identification and containment of C. auris are critical.
Article Details
Authors (5)
Woong Sik Jang
Young Lan Choe
Soo Young Yoon
Chae Seung Lim
Min-Chul Cho