Benchmarking the utility of dry-electrode electroencephalography for clinical trials
Abstract
Abstract This study investigated if new dry-electrode technologies for electroencephalography (EEG) can substantially lower patient and site burden in clinical trials while maintaining adequate data quality. We benchmarked three dry-electrode EEG devices against a standard EEG using typical clinical trial procedures and EEG tasks that are often used for biomarker purposes. We found that dry-electrode EEG can perform on par with standard EEG for a range of different applications. However, both the participant and technician acceptance varied strongly across devices. Dry-electrode EEG was able to only match the comfort of standard EEG at best but was faster and easier to work with. Consequently, dry-electrode EEG was ranked among the most and least preferred options. The quantitative performance of dry-electrode EEG varied strongly across different applications. For example, quantitative resting state EEG and P300 evoked activity were adequately captured by dry-electrode EEG. However, certain signal aspects, such as low frequency activity (< 6 Hz) and induced gamma activity (40–80 Hz) presented notable challenges for dry-electrode EEG. Our findings suggest that dry-electrode EEG can substantially improve clinical trial applications of EEG, if the device and its context of use are carefully matched.
Article Details
Authors (5)
Joseph Paillard
Philipp Bomatter
Laura Dubreuil-Vall
Jörg Felix Hipp
David Johannes Hawellek