Accuracy of infusion flow rates and bolus doses for portable infusion pump

R Ryo Sekiguchi M Michiko Kinoshita Y Yuki Maeda (Energy Chemistry Division, Energy Transformation Research Laboratory) K Katsuya Tanaka

Abstract

Abstract This study aimed to investigate the flow rate accuracy and bolus doses of four portable pumps with different mechanisms. We evaluated two electronic (COOPDECH Amy PCA [Amy] and CADD-Solis) and two elastomeric balloon-type pumps (COOPDECH Balloonjector and Rakurakufuser) under the following conditions: placement with an epidural catheter, operation on a shaking table, warming to 32 °C, and at room temperature (control, 25 °C). Amy maintained consistent flow rates across all conditions. The CADD-Solis also exhibited consistent flow rates, with minor yet statistically significant changes upon epidural catheter placement (− 5.0%, P = 0.005) and motion conditions (4.2%, P = 0.015). The Balloonjector and Rakurakufuser exhibited flow rate variations over time, and temperature increases significantly increased flow rates by 24.3% (P < 0.001) and 20.3% (P < 0.001), respectively. Bolus volume accuracy for the Amy, Balloonjector, and Rakurakufuser was not significantly affected under different conditions. The CADD-Solis showed a slight decrease in bolus volume with an epidural catheter (mean difference, 0.3 mL; P = 0.003). The electronic pumps maintained consistent flow rates across various conditions, whereas elastomeric balloon pumps exhibited variable rates influenced by time and temperature, increasing the risk of medication overdose. Bolus dosing accuracy was clinically satisfactory for all pump mechanisms.

Article Details

Volume / Issue Vol. 15, Issue 1
Published April 19, 2025
ISSN 2045-2322
Publisher Nature Portfolio

Journal Info

Scientific Reports

Nature Portfolio

ISSN: 2045-2322 Open Access Life Sciences

Authors (4)

R

Ryo Sekiguchi

M

Michiko Kinoshita

Y

Yuki Maeda

Energy Chemistry Division, Energy Transformation Research Laboratory

K

Katsuya Tanaka