Abstract 4369624: Unscrewing Sign: A Visual Indicator For Atrial Leadless Pacemaker Dislodgment

S Syed Hasan Raza Abidi (MedStar Washington Hospital Center, Washington, District of Columbia, United States) B Binaya Basyal (Medstar Washington Hospital Center, Washington, District of Columbia, United States) N Nebu Alexander (MedStar Washington Hospital Center, Washington, District of Columbia, United States) B Benjamin Szoch (Abbott, Sylmar, California, United States) C Cyrus Hadadi (MedStar Washington Hospital Center, Washington, District of Columbia, United States)

Abstract

Backround: The AVEIR DR i2i Study demonstrated safety of atrial leadless pacemakers with a dislodgment rate of only 1.7%. Data on factors influencing dislodgments however remains limited. Known contributors include suboptimal electrical mapping parameters, absence of current of injury during implantation, and excessive device manipulation during positioning. We present two cases of atrial leadless pacemaker dislodgment occurring in the absence of these established risk factors, each exhibiting a distinctive post-implantation imaging finding consistent with progressive 'unscrewing' of the pacemaker, visually characterized by anticlockwise movement of the Chevron. This radiographic pattern may serve as an early indicator for impending device dislodgment. Case Presentation: Case 1: 41-year-old female with symptomatic bradycardia, syncope and complete heart block underwent a dual chamber leadless pacemaker implantation. Electrical Mapping and current of injury was excellent. The device was implanted per protocol with one and half clockwise rotation. On postoperative day one, atrial loss of capture was noted. Chest X ray showed exaggerated device motion and revision was recommended. The pacemaker was removed using snaring and a new atrial leadless pacemaker was implanted. Upon reviewing the images from the first procedure, it was noticed that the chevron marker had rotated back (anticlockwise) around its axis leading to 'unscrewing' of the device. Case 2: 84-year-old male with a history of atrial flutter ablation, type 2 Diabetes, and hypothyroidism presented with symptomatic bradycardia and sinus pauses. Patient underwent atrial leadless pacemaker implantation. Electrical Mapping and current of injury was excellent. Next morning, atrial loss of capture was noted. An abdominal x-ray revealed that the device had dislodged into the pelvis. Interventional radiology was consulted who retrieved the device via snare. Upon reviewing the images, we noted that post-implantation, the chevron sign had rotated anticlockwise half a rotation. Discussion/Conclusion: Our cases did not have any of the known risk factors for dislodgment. The anticlockwise movement of the chevron sign may serve as an early indicator of dislodgment. Particular attention should be paid to the movement of the chevron post-procedure. Other associations such as hypertrophic myocardium, genetics, and valvular and rhythm abnormalities could play a role in unscrewing the pacemaker and need to be closely studied.

Article Details

Journal Circulation
Volume / Issue Vol. 152, Issue Suppl_3
Published November 04, 2025
ISSN 0009-7322
Publisher Lippincott Williams & Wilkins

Journal Info

Circulation

Lippincott Williams & Wilkins

ISSN: 0009-7322 Health Sciences

Authors (5)

S

Syed Hasan Raza Abidi

MedStar Washington Hospital Center, Washington, District of Columbia, United States

B

Binaya Basyal

Medstar Washington Hospital Center, Washington, District of Columbia, United States

N

Nebu Alexander

MedStar Washington Hospital Center, Washington, District of Columbia, United States

B

Benjamin Szoch

Abbott, Sylmar, California, United States

C

Cyrus Hadadi

MedStar Washington Hospital Center, Washington, District of Columbia, United States