Abstract 4364029: Biventricular Pressure-Volume Loop Assessment of Impella 5.5 in a Porcine Model of Myocardial Infarction
Abstract
Introduction: Microaxial flow pumps reduce mortality in myocardial infarction (MI) related cardiogenic shock (CS). The Impella 5.5 generates increased flow to provide circulatory support, and optimization relies on mechanistic understanding of pressure-volume (PV) relationships. Aim: Using a porcine model of acute MI in tandem with biventricular (BiV) PV loop conductance catheters, this study aimed to understand the impact of Impella 5.5 on BiV hemodynamics in MI CS. Methods: In 6 Domestic Yorkshire Pigs, an Impella 5.5 (J&J MedTech, NJ) was inserted percutaneously under fluoroscopic guidance via carotid or aortic access. An angioplasty balloon catheter was advanced past the first diagonal and inflated to above nominal pressure with blood flow obstruction for 90 minutes. BiV PV loop conductance catheters (Sigma-M, CD Leycom, Hengelo, The Netherlands) were inserted via femoral venous access. The left ventricle (LV) PV loop catheter is positioned via transeptal puncture using intracardiac echocardiography. BiV PV loops are generated for each level of Impella support from P1 to P7. Results: With increasing Impella 5.5 support, LV PV loops shifted to the left and right ventricular (RV) loops to the right. Without statistical significance, LV end diastolic volume (EDV) decreased by a mean of 12.3 ± 10.7% from baseline and RV EDV increased by 32 ± 40%. Native LV function decreased with increasing Impella 5.5 support, demonstrated by a reduction in stroke volume (SV) (35±13 ml to 27±14 ml; p<0.001), cardiac output (CO) derived from SV&heart rate (2.0±0.5 ml/min to 1.7±0.7 ml/min; p<0.001), and ejection fraction (EF) (37±7% to 29±12%; p<0.001). Less consistent trends were noted in peak positive dP/dt and end-systolic pressure volume relationship. Native RV function increased with higher levels of Impella 5.5 support, noted by increasing SV (27±24 ml to 41±28 ml; p=0.004), CO (1.6±1.2 ml/min to 2.3±1.4 ml/min; p=0.001), and EF (42±16% to 45±9%; p=0.04). Arterial elastance and ventriculo-arterial coupling increased in the LV from baseline to P7 (2.2±1.4 mmHg/ml to 2.8±1.0 mmHg/ml and 1.8±0.5 to 3.2±1.9, respectively; p<0.001) and decreased in the RV (1.3±0.7 mmHg/ml to 0.9±0.7 mmHg/ml (p<0.001) and 2.1±2.2 to 1.3±0.6 (p=0.004)). Conclusions: Impella 5.5 speed modulation highlights significant ventricular interdependence in the setting of acute MI. PV loop derived parameters here visualized parameters that may be utilized to improve optimization of hemodynamic support in CS.
Article Details
Authors (9)
Sara Inglis
Mayo Clinic, Rochester, Minnesota, United States
Serena Rahme
Mayo Clinic, Rochester, Minnesota, United States
Tyra Witt
Mayo Clinic, Rochester, Minnesota, United States
Mary Nagel
Mayo Clinic, Rochester, Minnesota, United States
Sarrafian Tiffany
Mayo Clinic, Rochester, Minnesota, United States
Katrina Tollefsrud
Mayo Clinic, Rochester, Minnesota, United States
suraj yalamuri
Mayo Clinic, Rochester, Minnesota, United States
Andrew Rosenbaum
Mayo Clinic, Rochester, Minnesota, United States
Atta Behfar