Abstract 4363387: Porcine Model of Biventricular Pressure-Volume Loop Assessment during Impella 5.5 Support

S Sara Inglis (Mayo Clinic, Rochester, Minnesota, United States) S Serena Rahme (Mayo Clinic, Rochester, Minnesota, United States) T Tyra Witt (Mayo Clinic, Rochester, Minnesota, United States) M Mary Nagel (Mayo Clinic, Rochester, Minnesota, United States) S Sarrafian Tiffany (Mayo Clinic, Rochester, Minnesota, United States) K Katrina Tollefsrud (Mayo Clinic, Rochester, Minnesota, United States) S suraj yalamuri (Mayo Clinic, Rochester, Minnesota, United States) A Andrew Rosenbaum (Mayo Clinic, Rochester, Minnesota, United States) A Atta Behfar

Abstract

Introduction: The Impella 5.5 is frequently employed to provide temporary mechanical circulatory support as bridge to recovery or durable therapy. Optimization of device management relies on understanding ventricular pressure-volume (PV) relationships. Aim: The aim of this study was to evaluate the biventricular (BiV) relationship with PV loop assessment during Impella 5.5 support in non-myopathic ventricles. Methods: Six domestic Yorkshire Pigs were used for this study. The Impella 5.5 (J&J MedTech, NJ) was inserted percutaneously under ultrasound and fluoroscopic guidance through either the carotid artery or aortic arch, using a 26Fr DrySeal Sheath (Gore, AZ). BiV PV loop conductance catheters (Sigma-M, CD Leycom, Hengelo, The Netherlands) were inserted via femoral venous access. The LV PV loop catheter is positioned via transeptal puncture and with intracardiac echocardiogram guidance. BiV PV loops are generated for each level of Impella 5.5 support. Results: With increasing levels of Impella 5.5 support, left ventricular (LV) PV loops shifted increasingly to the left and right ventricular (RV) loops to the right. Accordingly, LV end-diastolic volume (EDV) decreased (110±4 ml to 90±8 ml; p<0.001) and RV EDV increased (121±7 ml to 147±13 ml; p<0.001), from baseline to P9. Native LV systolic function decreased with increasing Impella 5.5 support, as demonstrated by a reduction in stroke volume (SV) (56±5 ml to 49±5 ml; p<0.001), cardiac output (CO) (3.4±0.8 ml/min to 2.0±0.3 ml/min; p<0.001), ejection fraction (EF) (51±4% to 44±6%; p<0.001), stroke work (SW) (4483±1023 ml*mmHg to 2383±719 ml*mmHg; p<0.001), and LV peak positive dP/dt (825 ± 184 mmHg/s to 523 ± 99 mmHg/s; p<0.001). Conversely, the native RV experienced preload-mediated increase in SV (65±7 ml to 86±20 ml; p<0.001), CO (3.7±0.8 ml/min to 4.3±0.9 ml/min; p<0.001), EF (53±4% to 58±11%; p<0.001), SW (918±143 ml*mmHg to 1529±365 ml*mmHg; p<0.001), and RV dP/dt (164±16 mmHg/s to 228±29 mmHg/s, p<0.001). Load independent contractility was assessed in the LV (1.64±0.33 mmHg/ml to 1.71±0.41 mmHg/ml; p=0.2) and RV (0.25±0.04 mmHg/ml to 0.29±0.05 mmHg/ml; p<0.001). Conclusions: This study highlights ventricular interdependence with Impella 5.5 support with opposing BiV trends. In a healthy heart, the RV can adequately compensate for increased preload offered by the device.

Article Details

Journal Circulation
Volume / Issue Vol. 152, Issue Suppl_3
Published November 04, 2025
ISSN 0009-7322
Publisher Lippincott Williams & Wilkins

Journal Info

Circulation

Lippincott Williams & Wilkins

ISSN: 0009-7322 Health Sciences

Authors (9)

S

Sara Inglis

Mayo Clinic, Rochester, Minnesota, United States

S

Serena Rahme

Mayo Clinic, Rochester, Minnesota, United States

T

Tyra Witt

Mayo Clinic, Rochester, Minnesota, United States

M

Mary Nagel

Mayo Clinic, Rochester, Minnesota, United States

S

Sarrafian Tiffany

Mayo Clinic, Rochester, Minnesota, United States

K

Katrina Tollefsrud

Mayo Clinic, Rochester, Minnesota, United States

S

suraj yalamuri

Mayo Clinic, Rochester, Minnesota, United States

A

Andrew Rosenbaum

Mayo Clinic, Rochester, Minnesota, United States

A

Atta Behfar