Abstract 4360977: Tissue-Transducer Contact Sensing Improves High-Intensity Ultrasound Catheter Ablation in Healthy Swine
Abstract
Background: High-intensity ultrasound (HIU) catheter ablation produces deeper lesions than radiofrequency and may serve as a new modality for interventricular septal (IVS) reduction in hypertrophic cardiomyopathy (HCM). A proprietary “contact sensing” algorithm was developed to confirm orientation of the side-facing transducer toward the IVS. Hypothesis: Contact sensing will improve HIU ablation efficacy, resulting in larger and more consistent IVS lesions. Methods: A total of 31 healthy swine underwent HIU catheter (Fig A) ablation via femoral venous access targeting the RV side of the IVS. In 13 of the 31 swine, continuous transducer-tissue contact was monitored to confirm tissue contact (Fig B) and optimize catheter positioning prior to ablation. Each swine underwent 8-16 HIU sonications. Animals were euthanized immediately post-ablation, and lesion presence, dimensions, volume, and location were assessed by gross pathology. Lesion metrics are reported as mean ± standard deviation. Group comparisons (with vs. without contact sensing) were performed using unpaired, two-tailed homoscedastic t-tests for continuous variables and chi-squared tests for categorical outcomes. Results: Contact sensing increased lesion volume from 657 ± 901 to 2071 ± 1230 mm 3 (p=0.001) and lesion depth from 7 ± 7 mm to 15 ± 3 mm (p=0.001; Fig C,D). Spline analysis of gross pathology showed that the lesion area on the LV side of the IVS increased from 22 ± 23% without sensing to 55 ± 19% with sensing (p=0.0004; Fig C,D). Transient atrio-ventricular block occurred in both groups (Fig D) and resolved with catheter repositioning. Sustained AVB occurred in one swine without contact sensing. Conclusion: HIU transducer-tissue contact sensing significantly improved HIU ablation efficacy, producing larger, deeper, and more optimally located lesions for HCM septal reduction. Survival studies will confirm the degree of septal reduction.
Article Details
Authors (8)
Craig Morris
University of Washington, Seattle, Washington, United States
Naveed Arang
University of Washington, Seattle, Washington, United States
Phoebe Dijour
Coridea, New York, New York, United States
Jeff Thiel
University of Washington, Seattle, Washington, United States
Stephanie Totten
University of Washington, Seattle, Washington, United States
Yak-Nam Wang
University of Washington, Seattle, Washington, United States
David Giraud
University of Washington, Seattle, Washington, United States
Babak Nazer
University of Washington, Seattle, Washington, United States