Abstract 4342332: A Non-Contacting Blood Flow Sensor for Assessing Pediatric Vascular Graft Patency

R Ruitong Chen T Trevor Hudson (University of Southern California, Los Angeles, California, United States) X Xuechun Wang (Department of Biological Sciences, Boler-Parseghian Center for Rare Diseases, Harper Cancer Research Institute, University of Notre Dame) J Jingjing Liang A Alanna Nowlen (University of Southern California, Los Angeles, California, United States) V Vishal Nigam (Seattle Children's Hospital, Seattle, Washington, United States) E Ellis Meng (University of Southern California, Los Angeles, California, United States)

Abstract

Introduction: For infants born with single ventricle physiology, the initial palliation involves the implantation of a systemic to pulmonary vascular graft, typically a modified Blalock-Taussig shunt (mBTS) or a Sano shunt. However, irregular blood flow through the life-sustaining shunt can be life-threatening, and state-of-the-art monitoring approaches are either invasive or indirect. Alternative approaches for periodic monitoring of blood flow in a nonclinical setting are needed. Methods: A novel contactless blood flow sensor was developed for integration on the shunt exterior for early detection of flow irregularities. This sensing strategy avoids device-related thrombosis by combining the capacitively-coupled contactless conductivity detection (C4D) principle with time-of-flight measurements to quantify blood flow. Five metal rings are positioned along the shunt exterior, each capacitively-coupled to the conductive blood. The first and last electrodes are used to impose a sinusoidal driving signal and voltage differences are measured between adjacent pairs among the three central electrodes. Time varying conductivity can be tracked and by cross-correlation of the two recordings, in which the downstream measurement is a time-delayed copy of the upstream one, and knowing the precise spacing of the pairs, flow rate can be computed. Results: Sensor prototypes consisted of platinum rings fitted on polytetrafluoroethylene tubing. Capacitive coupling of the rings to blood was evaluated first. Response to varying solution conductivities was then characterized, demonstrating a linear response with the highest sensitivity at lower conductivities. Following successful sensor calibration measurements in test fluid, ex vivo flow measurements using bovine blood were collected at 30 and 50 mL/min. Based on the electrode locations, the delay for the flow rates was expected to be 100 and 70 ms, respectively. Peaks were observed at 100 ms and 60 ms, corresponding to 30 mL/min (0% error) and 42.9 mL/min (14.2% error), respectively. Errors were comparable to current “gold-standard” devices. Conclusions: We successfully demonstrated the feasibility of non-contact flowing monitoring in a format suitable for pediatric vascular shunts. This low power approach can support a fully implantable device to provide early warning of inadequate blood flow and inform timely intervention to reduce morbidity and mortality in infants with single ventricle physiology.

Article Details

Journal Circulation
Volume / Issue Vol. 152, Issue Suppl_3
Published November 04, 2025
ISSN 0009-7322
Publisher Lippincott Williams & Wilkins

Journal Info

Circulation

Lippincott Williams & Wilkins

ISSN: 0009-7322 Health Sciences

Authors (7)

R

Ruitong Chen

T

Trevor Hudson

University of Southern California, Los Angeles, California, United States

X

Xuechun Wang

Department of Biological Sciences, Boler-Parseghian Center for Rare Diseases, Harper Cancer Research Institute, University of Notre Dame

J

Jingjing Liang

A

Alanna Nowlen

University of Southern California, Los Angeles, California, United States

V

Vishal Nigam

Seattle Children's Hospital, Seattle, Washington, United States

E

Ellis Meng

University of Southern California, Los Angeles, California, United States